The Gemini R is a black body source with which radiation pyrometers, commonly referred to as IR sensors, can be calibrated. With a temperature range of 50°C to 550°C, the temperature of the cavity can be set with 0.1°C resolution. The temperature of the furnace is set and indicated on a PID controller. An independent indicator is also provided for higher accuracy with a standard probe, which indicates the actual radiance temperature. Temperature variation over a 5-minute period, as viewed on the temperature controller, is
±0.2°C, and a radiation thermometer directed into the cavity detects similar variations.
The Gemini R's cavity's size is 2.5" x 6.3" deep (65mm x 160mm). Maximum temperature differences are ±1°C in the range of 100°C to 500°C. The Gemini R blackbody cavity is capable of accepting black body fixed points to achieve the highest level of a radiation measurement possible. Typical fixed-point uncertainties are ±0.25°C. Gallium, indium, tin and zinc cells are available.
Emissivity is greater than 0.995, and traceability may be established with a NIST certificate for the built-in indicator and/or supplied probe. The Gemini R can be supplied with RS232 or RS422 communications.
Model
|
976
|
Temperature Range
|
30°C to 550.0°C
|
Emissivity
|
Greater than 0.995
|
Stability
|
±0.1°C
|
Display Resolution
|
0.01°C to 99.99
0.1°C from 100 to 550
|
Heating Time
|
45 minutes
|
Aperture Diameter
|
65mm
|
Cavity Depth
|
160mm
|
PC Interface
|
included
|
Power
|
1000 Watts typical
|
Voltage
|
100- 130 or 208 -240 Vac 50/60Hz
|
Dimensions
|
H 310mm
W 265mm D 200mm
|
Weight
|
22lb
|
|
|
Options
|
|
Fixed Point Cell
|
431-03-00
|
Gallium Hockey Puck Cell
|
431-03-00
|
Indium Hockey Puck Cell
|
976-05-00A
|
Tin Hockey Puck Cell
|
976-05-00B
|
Zinc Hockey Puck Cell
|
976-05-00C
|
Orifice Plates 10, 20, 30, 40, 50mm
(Restricts Cavity Aperture)
|
976-01-05
|
Carrying Case
|
931-22-64
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Literature
|
|
|
|
|
Gemini R Datasheet
|
Gemini R Manual
|
|
|
|
|
|
|
|